Configuration of X-Ray Tube
Model: AT-42, Origin: India
High performance X-Ray vacuum tube compatible with C-Arm Machine
Both operable facilities of Radiography and Fluoroscopy
Longer life with Precious metallic but users friendly
Technical Data:
Tube type: Rotating anode, Dual focus
Anode heat storage capacity: 42KHU
Focal spots: Small focus : 0.6 mm
Large focus: 1.5 mm
Anode angle: 12 degree
Target Material : Tungsten
Filtration: Total Inherent Filtration is 0.8mm Al (added filtration 2.2 Al)
Collimator: 1 no. Single Leaf Parallel Shutter Collimator |
Configuration of X-Ray Tube
Model: AT-300, Origin: India
High performance X-Ray vacuum tube compatible with C-Arm Machine
Both operable facilities of Radiography and Fluoroscopy
Longer life with Precious metallic but users friendly
Technical Data:
Tube type: Rotating anode, Dual focus
Anode heat storage capacity: 300KHU
Focal spots: Small focus: 0.3 mm
Large focus: 0.6 mm
Anode angle: 10 degree
Cooling system: Oil cooling
Filtration: Total Inherent Filtration is 0.7mm Al (added filtration 2.3 Al)
Collimator: 1 no. Single Leaf Parallel Shutter Collimator |